Factorycalibration Reset - Tintometer Lovibond MD200 Instrucciones

Ocultar thumbs Ver también para Lovibond MD200:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

GB
Calibration Mode
Store
Date
Cal
RESULT
Time
Cal
Zero
Mode
Test
On
Off
SEL
CAL
SEL
cAL
Mode
SEL
CAL
On
Off
MD200_1a 08/2011
Factory calibration reset
Resetting the user calibration to the original factory calibration will
reset all methods and ranges.
A user calibrated method is indicated by an arrow while the test result
is displayed.
To reset the calibration press both the [MODE] and [ZERO/TEST]key
and hold.
Switch the unit on using the [ON/OFF]key.
Releasethe [MODE] and [ZERO/TEST]keysafter approx. 1 second.
The following messageswill appear in turn on the display:
The factory setting is active.
(SELstands for Select)
or:
Calibration has been set by the user.
(If the user calibration is to be retained, switch the unit off using the
[ON/OFF] key).
Calibration is reset to the factory setting by pressing the [MODE] key.
The following messageswill appear in turn on the display:
Switch the unit off using the [ON/OFF]key.
23
Tabla de contenido
loading

Tabla de contenido