NOTE 1: at 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: these guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Oxystart
level above, the Oxystart
performance is observed, additional measures may be necessary, such as reorienting or relocating the
Oxystart
Pulse Oximeter.
®
b) Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
52
Oxystart® - EN
Pulse Oximeter is used exceeds the applicable RF compliance
®
Pulse Oximeter should be observed to verify normal operation. If abnormal
®
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a) should be less than the compliance level in
each frequency range b)
Interference may occur in the vicinity of equip-
ment marked with the following symbol: