Power Consumption: less than 30mA.
3.
4.
Resolution: 1% for SpO
5.
Measurement Accuracy: ±2% for values from 70%-100% SpO
70%. ±2 bpm or ±2% (select the highest) for Pulse Rate.
6.
Measurement Performance under Poor Filling Condition: SpO
when the pulse-filling ratio is 0.4%. SpO
the highest).
Resistance to surrounding light: The deviation between the value measured under conditions with man-
7.
made lighting or indoor natural lighting and that in a darkroom is less than ±1%.
8.
The device is equipped with a function switch. The Oximeter powers off if no finger is detected in the
Oximeter for 5
seconds.
9.
Optical Sensor
Red light (wavelength of 660nm, 6.65mW)
Infrared (wavelength of 880nm, 6.75mW)

5. Accessories

› One strap;
› Two batteries (optional);
› One User Manual.
and 1 bpm for Pulse Rate.
2
accuracy is ±4%, pulse rate accuracy is ±2 bpm or ±2% (select
2
, and negligible for values not exceeding
2
and pulse rate can be shown correctly
2
Oxystart® - EN
39