MC01 R / RL - Instructions for use - EN
7.7 Sterilisation
Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060/ISO 17665-1
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
Before each sterilisation cycle, service the medical device with Faro care products.
CAUTION
Contact corrosion due to moisture.
Damage to product.
Immediately remove the product from the steamsteriliser after the sterilisation cycle!
The Faro medical device has a maximum temperature resistance up to138° (280.4 °F).
Depending on the available autoclave, select a suitable procedure from the following sterilisation processes:
▪ Autoclave with three times initial vacuum:
– at least 3 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4°F)
▪ Autoclave using the gravitation method:
– at least 10 minutes at 134 °C -1 °C / +4 °C (273 °F -1.6 °F / +7.4°F) or alternatively
– at least 60 minutes at 121 °C -1 °C / +4 °C (250 °F -1.6 °F / +7.4°F)
Use according to the manufacturer's Instructions for Use.
56