Faro MC01 R Manual De Instrucciones De Uso Y Cuidado página 54

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

MC01 R / RL - Instructions for use - EN
7.3 Disinfection
CAUTION
Malfunctioning from using a disinfectant bath or disinfectant containing chlorine.
Defects in the product.
 Only disinfect in a thermodisinfector or manually
7.3.1 Disinfection: Manual disinfection – external
Faro recommends the following products based on material compatibility.
The microbiological efficacy must be ensured by the disinfectant manufacturer.
▪ Mikrozid AF Liquid made by Schülke & Mayr
▪ FD 322 made by Dürr
▪ CaviCide made by Metrex
Consumables required:
▪ Cloths for wiping off the medical device.
 Spray the disinfectant on a cloth, then thoroughly wipe down the medical device and leave the disinfectant to soak in according to the instructions from the disinfectant
manufacturer.
 Follow the instructions for use of the disinfectant.
52