4 Leave all other fields blank.
5 Select Done. The guideline is blue when training mode is active.
To exit training mode
1 Press the PATIENT key.
2 Select New/End.
3 Type any name other than @AXOTRACK in the Last Name field.
4 Select Done.
Measurements and calculations
The distance between the tip of the graphic overlay (center of needle graphic ellipse) and the sonographic tip
is within 2 mm plus 1% of the depth of the needle tip. Only calculations related to 2D, M Mode, and Color apply
to the P11x transducer.
Measurement accuracy
The measurements provided by the system are verified on a static bench model and do not account for
acoustic anomalies of the body. The 2D linear distance measurement results are displayed in centimeters with
two places past the decimal point. The linear distance measurement components for the P11x transducer
have the accuracy of +/- 0.4 cm.