2 Protector
WARNING!
Risk of injury for patient or hospital staff due to incorrect installation of protector:
►
Only people who are familiar with this user manual are allowed to perform installation of protector,
removal of protector or other manipulation with installed protector! Manufacturer is not responsible for
possible injuries of patient or other persons caused by non-observance of the instructions
in this user manual!
►
It is not allowed to perform installation of protector, removal of protector or other manipulation with
installed protector when there is a patient on the bed!
►
After installation or removal, ensure that each safety screw is tightened with torque 5+2 Nm!
►
Before use the protector must be installed on the bed correctly according to this user manual!
►
Protector is not compatible with Infusion stand installed in the corner of the Tom 2 bed!
WARNING!
Risk of injury for patient or hospital staff due to incorrect use of protector:
►
Hospital staff is responsible for assessing if the physical and psychological state of the patient is in
accordance with use of the protector!
►
Hospital staff is responsible for assessing if the patient may be left unattended on the bed with installed
protector!
►
Hospital staff is responsible for assessing whether bedclothes, toys or other objects of larger
dimensions increase the risk of patient´s climbing over the siderails or bed ends!
►
Use of protector is not allowed in emergency cases or crisis situations where the patient or hospital staff
could be injured!
►
USE OF PROTECTOR IS NOT ALLOWED IF SIDERAILS ARE NOT UPRIGHT IN THE HIGHEST POSITION!
►
Use of protector is not allowed if bed ends are not correctly locked or are removed from the bed!
►
During use of the protector, it is not allowed to adjust backrest and mattress platform should be in the
horizontal position!
2.1 Intended use
The protector is intended for reduction of probability that the patient will leave the bed.
Protector is not 100% protection against patient leaving the bed.
2.2 Mechanism of effect
The protector raises the height of both the siderails and the bed ends based on its location. The shape and rotating moti-
on of the protector make it more difficult for the patient to grasp the edges of the bed and climb out.
2.3 Location
The protector is installed around the perimeter of the bed (above siderails and bed ends). It is firmly fixed on each of the 4
upper corners of the bed.
2.4 Instructions for installation
It is recommended that 2 people install the protector.
1)
Connect all 4 protector rollers so that pins of shorter rollers are inserted into the bushings of longer rollers
(position 1 on both pictures below).
2)
Insert pins of the assembled protector into bushings on upper corners of the bed (position 2 on both pictures
below).
3)
Fix the protector with safety screws on the 4 upper corners of the bed (position 3 on second picture below).
4)
After each installation ensure that safety screws on the 4 upper corners of the bed are tightened and
the protector is firmly fixed with them. Use an allen key (size 5 mm) to tighten screws with torque 5+2 Nm.
D9U001KRB-01ML_01