and it
cannot
be
changed)
2.2 Zero setting key range control <
2%
(On)
(If it is approved, its
default value is ON
and it
cannot
be
changed)
2.3 Zero setting depends on stability
(On)
(If it is approved, its
default value is ON
and it
cannot
be
changed)
2.4 Automatic zero setting
(On)
If
the
weight
indication is negative
for
more
than
5
seconds when the
equilibrium indicator
is on, zero will be set
automatically.
2.5 Automatic zero tracking
(On)
If it is ON
:
Automatic
zero
tracking
is
done
when the display is
zero and stable.
If
it
is
OFF
:Automatic
zero tracking will not
be done.
NOTE
:
If
there
is
tare,
automatic
zero
tracking program is
not going to run.
2.6 Zero Setting via interface
(Off)
Not in this type
2.7 Silo Mode Enable/Disable
(Off)
If it is ON :Silo Mode
is enabled and initial
zero setting control
is not done.
If it is OFF :Silo
Mode is disabled.
3. Total Memory, Unit Price:
3.1 Adding without showing total
memory
(On)
If it is ON: Total
cannot be seen after
M+.
Weighing
seen.
If it is OFF: Total can
be seen until the
weighing
empty, after M+
3.2 Temporary unit price
(On)
Not in this type.
3.3
Adding by M+ is depends on
stability
(On)
(If it is approved, its
default value is ON
and it
cannot
changed.)
4. Limitation:
4.1 Limitation Lower Limit
The beep of buzzer
is started at lower
limit value.
4.2 Limitation Upper Limit
The beep of buzzer
is stopped at upper
limit value.
- 15 -
4.3 Limitation On/Off
(On)
4.4 Stability due to PC packet sending
lower limit
(1)
4.5 Stability due to PC packet sending
upper limit
(1)
4.6 Stability due to PC packet sending
interval tolerance
(20)
4.7 Stability due to PC packet sending
is
minimum equilibrum duration
10 -> 1 second
4.8 Stability due to PC packet sending
plate
is
On/Off
(On)
5. LCD / Time Settings:
5.1
be
5.2
If this specification is
ON,
limitation
is
done.
6. Power management settings:
(10)
LCD contrast
(7)
Can
be
adjusted
between 1 and 9
Backlight brightness
(7)
Can
be
adjusted
between 1 and 9
- 16 -
5.3
Date
(Optional)
First
two
digits
:
day
Middle
two
digits:
month
Last two digits:
year
5.4
Time
(Optional)
First
two
digits:
hour
Last two digits:
minute
6.1 Empty weighing plate special
display
(On)
If it is ON: there will
be a flashing point
on display instead of
weight value if scale
is
powered
by
battery and empty
during 30 seconds.
(in a scale with LCD
display "
"
------
is shown)
If it is OFF: does not
work.
6.2 Empty weighing plate special
display
(On)
Manages
Empty
Weighing
Plate
Special
Display
mode
and
other
power management
modes.
If it is ON: Power
management is done
for all modes.
If it is OFF:
No
power management
is done and "
----
" is not shown
--
also.