Disinfection; Disassembly - Drive Voyager Pro 18018 Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Disinfection

NOTE: The mesh nebulizer should ONLY be cleaned with water.
.First clean the nebulizer container and the accessories as described in the
cleaning section. You can then continue with disinfection.
.Pour 8 ml ethyl alcohol (75% v/v) into the medicine container. Close
the container. Leave the alcohol in the container for at least 10 minutes.
.For better disinfection, shake the device gently from time to time.
Be sure the lid is properly fixed to avoid spilling the solution.
.Afterwards, pour the ethyl alcohol out of the container.
.Repeat this process, using water this time.
.As an alternative to ethyl alcohol, a solution of water and vinegar may
be used. Be sure to rinse with clean water and thoroughly dry.

Disassembly

. To continue with cleaning
its individual components,
disassemble the nebulizer.
. Remove the medicine container
and mesh nebulizer by pressing
both brackets inwards and
drawing the device out.
17

Capítulos

Tabla de contenido
loading

Tabla de contenido