Troubleshooting And Display Symbols; Error Message - Veridian Healthcare 09-332 Instrucciones

Termómetro temple touch
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

TROUBLESHOOTING / DISPLAY SYMBOLS
The thermometer has been designed to display feedback-warnings to ensure
proper usage and accurate temperature measurement. When an error appears, the
device will "beep" 5 times, and then the thermometer will turn off automatically.
In the event of a malfunction or incorrect temperature measurement, the
following warnings will be displayed:
ERROR
MESSAGE
16 • ENGLISH
PROBLEM/CAUSE
The measured temperature is
below 95°F (35°C)
The measured temperature is
above 107.6°F (42°C)
The thermometer is too cold or too
hot to operate
Temperature did not register be-
cause of poor contact, patient
movement or poor placement
Unit malfunctioned
Low battery - Battery icon displays
continuously
Drained battery - Battery icon is the
only icon the display and unit turns
off
SOLUTION
1. Turn the unit OFF and then ON again
2. Retake temperature
1. Turn the unit OFF and then ON again
2. Retake temperature
1. Verify that the room temperature
within product range 60.8°F-104°F (16° C
- 40° C)
2. Retake temperature
Turn the unit OFF, wait 2 minutes, turn it
ON again and retake temperature
Do not use the thermometer – contact
Customer Care
Battery is low; enough power remains for
approximately 10 measurements
Battery is completely expired and must
be replaced

Capítulos

Tabla de contenido
loading

Tabla de contenido