NOTES ON PROGRAMMING
Menu
Setting
9.Offset
probe
ESC
Offset calibr.
Probe bottom gap
Bottom deadstock
Default values
10.Trim
xxx.x cm
height
Calibrate:No
Calibrate:Yes
11.Trim
xxx.xxx L
volume
Calibrate:No
Calibrate:Yes
12.Unit
L
m³
%
m
kg
IG
UG
t
mbar
kPa
Part No. 28 100 57 e
®
SmartBox
1 / SmartBox
Description
Adjusting:
• Probe zero point, electric
• Position / Distance from base
• Unusable capacity that is not to be displayed
Exit the menu
New measurement of probe zero point (electric)
Lift probe out of the liquid beforehand.
Distance:
normal reference is x = 0 cm, max = 99 cm
Suction position:
Normal reference is 0 cm = complete capacity.
y > 0 cm means corresponding unusable
capacity.
Reset all values to default 00.
Entry option for the reference height for the 2-
point measurement, for other probe measurement
range or for an unknown density. Subtract 1.0 cm
from the actual measured level and enter this
value.
If activated (Yes), the display in menu steps
"1.Measure probe" and "2.Liquid" is then "by
comparison".
If this is entered with an almost empty tank, it is
recommended that you make a correction the
next time it is filled.
Readjust the level display with the [+] / [-] button
See menu step "10.Trim height"
Default settings L
liter:
cubic meters:
percent:
meter:
kilogram:
imperial Gallon:
US liquid gallon:
ton:
millibar:
kilopascals:
®
2 / SmartBox
x cm;
y cm
999900 L
2.50 m³
99.50 %
2.50 m
999900 kg
219750 IG
263900 UG
2.50 t
500 mbar
50 kPa
®
3
11 / 16