The user can either calibrate the fuel gauge and charge the battery, or just charge the
battery.
Calibration takes longer than charging and it may not be convenient to go through the
calibration cycle.
To recalibrate the fuel gauge, press the button on the front of the charger.
The charger will automatically begin to charge the battery if the button is not pressed.
The blue LED will flash to indicate that the battery is undergoing the recalibration cycle.
During calibration the discharge resistors will be cooled by the fan. Removing the
battery, or pressing the calibration button again will re-start the process from the
beginning.
At the end of this procedure the blue LED will stay constant indicating a fully calibrated
fuel gauge. Warm environments can cause calibration failure - keep the charger away
from direct sunlight or heat sources.
7.

REPLACEMENT PARTS

For replacement parts, additional or optional equipment :
Please call Customer Service @ 877-677-2832
8.

USER SERVICE

The LED Headlight has no user or field serviceable components. It can only be serviced
at the factory by factory trained technicians.
System Function and Error LEDs are visible on the back side of the Battery holster
(side with the belt clip) There is one 'red' and one 'green' LED indicator lights.
When the unit is turned on:
The lit 'Green' LED signifies the system is powered up and will stay on constantly.
The lit 'Red' LED signifies the system requires service. Turn off the headlight and
remove the battery. Contact Customer Service.
9.
LIMITED WARRANTY
Your LED Headlight and battery holster carry a three year warranty from the date of
shipment on workmanship and all defects of material.
Your batteries and charger carry a six month warranty from the date of shipment on
workmanship and all defects of material. Should your product prove to have such
defects
ST TECHNOLOGIES
charge. Please contact
You should carefully pack unit in a sturdy carton and ship it to the factory. Please
include a note describing the defects, your name, telephone number and a return
®
LIT-171 CUDA
SURGICAL
(English)
®
will repair or replace the product or component part without
®
for return authorization documentation.
ST TECHNOLOGIES
Rev. D
Date of Revision:3/28/13
Page 8 of 49