Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
Recommended separation distances between portable and mobile RF communications equi-
pment and the ME EQUIPMENT or ME SYSTEM
The device is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the device can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the tran-
smitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
IEC 60601 test
level
contact ±8 Kv
air ±2, ±4, ±8,
±15 kv
30 A/m
50 Hz or 60Hz
Separation distance according to frequency of transmitter m
150KHz to 80MHz
√P
d = 1.2
N/A
N/A
N/A
N/A
N/A
Compliance
Electromagnetic environment -
level
guidance
contact ±8 Kv
Floors should be wood, concrete or
air ±2, ±4, ±8,
ceramic tile.
±15 Kv
If floors are covered with synthetic ma-
terial, the relative humidity should be
at least 30 %.
30 A/m
Power
50 Hz or 60Hz
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
80MHz to 800MHz
√P
d = 1.2
0.12
0.38
1.2
3.8
12
frequency
magnetic
800MHz to 2.5GHz
√P
d = 2.3
0.23
0.73
2.3
7.3
23
40
fields