12. If test result is above 600 mg/dL for Glucose, the screen dis-
plays HI.
If test result is below 20 mg/dL for Glucose, the screen dis-
plays LO.
NOTE: The monitor will time out after 2 minutes of non-use or if
the strip is removed. The keys are disabled until a strip is inserted.
Results and marking status are saved if the monitor times out, the
strip is removed, or the monitor is turned off.
NOTE: Lancets are for one-time use only. Use a new, sterile lancet
each time you test. Test different areas on your fingertips to avoid
developing calluses. Remove the used lancet from the lancing de-
vice. Follow your local disposal regulations where applicable.
WARNING: Your lancing device is for your personal use only. DO
NOT share with others. Sharing the lancing device or lancets can
transmit serious, even grave infections. To avoid accidental sticks,
do not store used lancets in the device after testing or arm the lanc-
ing device with a new sterile lancet unless ready to use.
NOTE: Do not press the test strip directly against the skin. Touch
the test strip gently to the blood drop.
20
Nova Max Plus Monitor OG PN46171D1 EN.indd 20
Running a Glucose Test
9/5/12 3:22 PM