Running Glucose Control Solution

Important Information for Glucose Control Solution
• Use only Nova Max Glucose Control Solutions for the test.
• Check the expiration date on the control solution vial. Do not
use control solution past the expiration date or you may get
inaccurate results.
• Store only for 3 months after first opening. When you open
a new vial of control solution, count forward 3 months and
write that date on the label of the control solution vial. Dis-
card any remaining solution after the date you have written
on the vial.
• Store the control solution tightly closed at room temperature
below 86°F (30°C). Do not refrigerate or freeze.
• Shake the control solution well before using.
Caution: The Nova Max Glucose Control Solution range printed on
the glucose test strip vial is for control solution only. It is used to test
the performance of the monitor and test strip. It is not a recom-
mended range for your blood glucose level.
If your control solution test results continue to fall outside the
range printed on the test strip vial:
• The Nova Max Plus Monitor may not be working properly.
• Do not use the monitor to test your blood.
• Inside the USA, call Customer Service at 1-800-681-7390.
Outside the USA, contact your Nova supplier.