Specification
Light source
Measurement range
Measurement interval 150 seconds
Accuracy
Precision
Limit of detection
Limit of quantitation
Calibration
Reagent usage
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.

2.1 Safety information

The manufacturer is not responsible for any damages due to misapplication or misuse of this
product including, without limitation, direct, incidental and consequential damages, and disclaims
such damages to the full extent permitted under applicable law. The user is solely responsible to
identify critical application risks and install appropriate mechanisms to protect processes during a
possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.

2.1.1 Use of hazard information

Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or
serious injury.
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death
or serious injury.
Indicates a potentially hazardous situation that may result in minor or moderate injury.
4 English
Table 3 Measurement specifications
Details
LED, measurement at 510 nm; 1 cm light pathlength
0.03–10 mg/L free or total residual chlorine as Cl
±5% or ±0.04 mg/L from 0 to 5 mg/L (the larger value) as Cl
±10% from 5 to 10 mg/L as Cl
±5% or ±0.01 mg/L (the larger value) as Cl
0.03 mg/L as Cl
2
0.07 mg/L
Factory calibration
Optional: 2-point user calibration with calibration standards in Calibration
Verification Kit
0.5 L of buffer solution and 0.5 L of indicator solution in 31 days