10 SERVICE AND MAINTENANCE
INFORMATION
The QuikRead go instrument has been designed to
be as user-friendly as possible, with no need for reg-
ular maintenance. In the event of any repair require-
ments, please contact your local supplier.
Cleaning the instrument
Periodically clean the exterior of the instrument us-
ing a lint-free cloth dampened with water. Pay spe-
cial attention to cleaning the display. Take care that
no liquid comes in contact with the edges of the
display, the measurement well, or the connectors.
If necessary, a mild detergent may be used. Do not
use organic solvents or corrosive substances. Spills
of potentially infectious material should be wiped
off immediately with absorbent paper tissue and
the contaminated areas swabbed with a standard
disinfectant or 70% ethyl alcohol. Materials used to
clean spills, including gloves, should be disposed of
as hazardous biowaste.
Software updates
New software can be uploaded to the instrument via
a USB drive, in line with the instructions presented
on the screen. Ask your local supplier for more in-
formation.
32
SERVICE AND MAINTENANCE INFORMATION
Changing the clock battery
The instrument has a battery that supplies power
to the internal clock. If the clock battery runs out,
a warning is displayed. The clock battery can be re-
placed with a battery of the same type (CR 2032 3V).
1.
Turn off the instrument (if it is on).
2.
Unplug the power-supply cable.
3.
Place the instrument on its side on a table.
4.
Open the battery-compartment cover.
5.
If a battery pack is inserted, unplug the battery-
pack connector from it, and remove the battery
pack.
6.
Remove the clock battery from the battery-
holder.
7.
Place a new clock battery (type CR 2032 3V) in
the battery-holder, text side down.
8.
If a battery pack is being used, reattach the bat-
tery-pack connector to the battery pack, press
the battery pack back into place, and ensure
that it is positioned properly. Close the battery-
compartment cover.
9.
Place the instrument back in an upright position
and plug in the power-supply cable.
10. Start the instrument by pressing the Power but-
ton.
11. Adjust the date and time (Settings > Personal
settings > Measurement flow > Maintenance >
Date and Time).