Guidance and Manufacturer's Declaration - Electromagnetic Immunity
(continued):
: Field strength from fixed transmitters, such as base stations for radio (cellular/cordless)
a
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Inogen One®
G2 Oxygen Concentrator is used exceeds the applicable RF compliance level above, the
Inogen One® G2 Oxygen Concentrator should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the device.
: Over the frequency range 150 kHz to 80 MHz, the field strengths should be less
b
than 3V/m.