HEINE
mPack mini
®
Please read and follow these instructions for use and keep them for future
reference.
Intended Use
The HEINE
mPack mini rechargeable battery is intended for use with HEINE medical
®
devices exclusively.
It may be used by professional medical staff and in professional healthcare institutions only.
For U.S. only:
Federal law restricts this device to sale by or on the order of a Physician or
Practitioner!s
Warnings and Safety Information
CAUTION! Indicates potential hazardous situations. Ignoring the corresponding
instructions may lead to dangerous situations of mild to moderate extent.
(Background color yellow; foreground color black).
NOTE! Note indicates valuable advice in terms of installation, operation,
maintenance or repair. Notes are important, but not related to hazardous situations.
Product overview
1 Micro-USB connector
2 Socket
3 Controller
4 Status indicator
Setting up
Charge the HEINE
Charging
Use a USB cable to connect the USB connector (1) of the mPack mini to a USB power
supply (e.g. HEINE E4-USB, PC or Laptop). At the start of the charging process, all three
illuminating segments of the status indicator (4) light up briefly. During the charging
process, the segments light up progressively to indicate the charging status. As soon
as all segments are on permanently, the mPack mini is fully charged. The indicator is
switched off after a short period of time.
Installation
Plug the supply cable of the HEINE instrument you wish to operate with the mPack mini
into socket (2).
Operation
Rotate the controller (3) to change the brightness of the connected instrument. To switch
the mPack mini off, rotate the controller counterclockwise until you hear a „click" .
When the mPack mini is switched on and an instrument is connected to it, the status
indicator indicates the charging status as follows:
- All segments are on permanently: fully charged
- Lower two segments are on permanently: sufficiently charged
- Lower segment (orange) is on permanently: almost discharged
- Lower segment (orange) flashes: strongly discharged, please recharge!
If no instrument is connected to the mPack mini when the mPack mini is switched on, the
status indicator lights up for approximately 1 second and is then switched off.