IMPORTANT:
performing Steps 1-4 in the Quick Start (pg. 5).
1. Press the power button
PosiTest HHD is now in Standby Mode (pg. 1).
2. Press the
the
or
-
+
3. Select the preferred Audio and Strobe Alarm settings (pg. 10).
4. Adjust the test voltage (pg. 7).
5. To activate the high voltage output and enable Detection
Mode, press and hold the power button
hold the safety trigger. The lightning bolt icon will appear at
the bottom of the display and the PosiTest HHD will show the
test voltage being generated.
6. Continue holding the safety trigger and release the power
button
.
7. Move the electrode over the coated surface at a travel speed
not to exceed 0.3m/second (1ft/second) to locate defects. If a
holiday is detected, the PosiTest HHD will activate the strobe
and/or the audio alarm (depending upon setup).
8. Release the safety trigger to deactivate the high voltage output
and return the PosiTest HHD to Standby Mode (pg. 1).
Instrument Operation
The PosiTest HHD is ready for testing only after
button to access the menu. Navigate using
buttons. Press
to power-up the detector. The
to select.
9
and activate and