Instructions
CAUTION: Prior to using this DermLite,
read these instructions and keep them in
a safe place for future reference. United
States Federal Law restricts this device to
sale by or on the order of a physician. This
DermLite is designed for external examina-
tion only.
DermLite Foto X is a polarized and non-
polarized dermatoscope designed for use
with a wide range of cameras and camera-
equipped mobile devices.
Directions for Use
CAUTION: Do not look directly into the
bright LED light. Patients must close their
eyes during facial examinations.
IMPORTANT: Before initial use, please
charge the internal lithium ion battery for a
minimum of 4 hours using the supplied USB
charging cable. See "Battery & Charging"
for details.
CAUTION: If your DermLite is used for im-
mersion fluid dermoscopy, place the fluid
only on the skin and then place the unit on
top of the fluid. Do not allow fluid to contact
the Charging Port (CP).
Foto X may be attached to many cameras or
smartphones via either a magnetic or purely
threaded connection. For use with Mag-
netiConnect™ adapters (sold separately),
screw the Ring Adapter (RA) onto the 37 mm
thread (37). To create a threaded connection
with your camera, remove the Ring Adapter
(RA) and thread the Foto X onto a compatible
camera or threaded camera adapter*.
Place the unit onto the skin, then push
the Power button (P) to activate the light.
Tapping the power button toggles between
polarized and non-polarized light. Move be-
tween four brightness levels via the Bright-
ness Buttons (BB). To adjust focus, turn the
Spacer (S) as needed; aligning the Focus
Marks (FM) creates the ideal focal distance
for most photography.
To turn off your DermLite, push and hold the
Power button (P) for about one second; the
unit turns off automatically after 3 minutes.
*Note that many cameras require the use of an
additional adapter made by the camera manu-
facturer or a photography supply company.
Removing the Spacer
The Spacer (S) may be removed by turning it
counter-clockwise past a slight resistance.
When reattaching, make sure its three small
protrusions match up with the Bayonet Slot
(BS).
Removing the Faceplate
The Faceplate (FP) may simply be pulled off
the Spacer (S) by grabbing it at the fingernail
notches. To reattach, simply snap it back in
place.
Infection Control and Epidemiology Caps
(IceCap™)
To support your infection control measures,
5 disposable IceCaps (IC) are supplied with
this device. To use, snap an IceCap over the
Spacer (S) with the Faceplate (FP) installed.
To remove and dispose, simply pull it off.
Battery & Charging
CAUTION: Your DermLite uses a special lith-
ium battery, which can only be purchased
from 3Gen or from an authorized 3Gen
dealer. Do not under any circumstances use
another battery other than the one designed
for this unit. Doing so could result in dam-
age to the unit.
This device is equipped with a four-level
Charge Indicator (CI). When the unit is
turned on and fully charged, all four LEDs
light up. When less than 75% of battery life
remains, three LEDs are illuminated. If more
than half the battery is drained, two LEDs
are on, while a single LED indicates that less
than 25% of battery capacity remains.
To charge your DermLite, plug the (USB-C)
end of the supplied cable into to the Charg-
ing Port (CP) and the other (USB-A) end
into any USB port or charger (not included).
When fully charged, the device will allow ap-
proximately 1 to over 5 hours of continuous
operation depending on brightness level.
After years of use, you may wish to replace
the battery, which is only available from
3Gen or an authorized 3Gen dealer.
To change the battery, contact 3Gen or your
local 3Gen dealer.
Troubleshooting
Please check www.dermlite.com for the
most current troubleshooting information. If
your DermLite requires servicing, visit www.
dermlite.com/service or contact your local
3Gen dealer.
Care and Maintenance
Your DermLite is designed for trouble-free
operation. Never attempt to open the device
for any reason.
Cleaning and Sterilization
The exterior of your DermLite, except the
optical parts, may be wiped clean with
isopropyl alcohol (70% vol.). The lens and
the polarizing filters should be treated as
high quality photographic equipment and
should be cleaned with standard lens clean-
ing equipment and protected from harmful
chemicals. Do not use abrasive material on
any part of the equipment or immerse the
device in liquid. Do not autoclave.
The DermLite Foto X set includes:
Foto X scope (24 polarized & 8 non-polarized
LEDs, 30mm 8x lens), USB-A to USB-C
charging cable, 5 IceCaps.
Warranty
5 years for parts and labor. Serial number
and company information are located on the
side of the device underneath the Spacer (S).
English