1.4.6.1.
Automatic Tare
The IND231 can be configured so that tare is taken automatically (auto tare) after the weight on the
scale exceeds a programmed tare threshold weight. Auto tare can be configured in setup as
enabled or disabled. When auto tare is enabled, the display changes to a zero net weight indication
after the weight exceeds the threshold value. Auto tare operations involve:
1.4.6.2.
Clearing Tare
Tare values can be cleared manually or automatically.
1.4.6.2.1.
Manual Clear
Manually clear tare values by pressing the CLEAR key
IND231/IND236 is in the net mode and has completed the weighing operation. Motion on the scale
will not impact a manual clear.
If configured in setup, pressing the ZERO scale function key
zero command.
1.4.6.2.2.
Auto Clear Tare
The IND231/IND236 can be configured to clear tare automatically when the weight returns to zero
point. Once the tare is cleared, the display returns to the gross weighing mode.
30100541
|
00
|
08/2013
Auto tare
When weight on the scale platform exceeds the tare threshold value,
threshold Weight
the terminal automatically tares.
Auto tare reset
The reset threshold weight must be less than the tare threshold weight.
Threshold Weight
When the weight on the scale platform falls below the reset threshold
value, such as when a load has been removed, the terminal
automatically resets the auto tare trigger.
on the keypad when the
will first clear the tare, then issue a
METTLER TOLEDO IND231/IND236 Quick Guide