Note: If Meter does not beep or begin countdown
soon after placing Sample Tip to Control
drop, discard Strip. Repeat test with new
Test Strip. If problem persists, see
Troubleshooting.
11. Compare result to Control Level
range printed on Test Strip vial
label:
If result is in range, System
can be used for testing blood.
If result does not fall within
range, repeat test using a new
Strip.
Caution! If Control Test result continues to be
outside range, System should not be
used for testing blood. Call for
assistance.
12. Remove Test Strip from Meter and discard.
Meter turns off.
13. Wipe Control drop off of vial cap with a clean,
wet tissue and dry thoroughly.