FLOW/INFUSION RATE:
1-295 mL/hr in 1 mL increments
DOSE LIMIT:
0-9995 mL in 5 mL increments
ACCURACY:
±10% of selected flow rate
OCCLUSION PRESSURE:
Does not exceed 15 p.s.i.
BATTERY:
• Sealed lead acid
• Rechargeable
• Automatically recharges when connected to
an AC outlet
• Recharge time is 12 hours from complete
discharge to full recharge
• Operating time is 8 hours at 100 mL/hr
(Charge unit fully every 30 days)
The backup battery is designed to provide
temporary power to the pump if AC outlet
power is interrupted. Frequent, extended
operation of the pump on the backup battery
will shorten battery life.
DIMENSIONS:
7.4" High x 9.5" Wide x 5.3" Deep
WEIGHT:
5.7 Pounds
CASE MATERIAL:
Fire Resistant Plastic
POWER REQUIREMENT:
120V ~, 187mA, 60Hz
Serial Number less than 700000:
Protected Against Vertically Falling Water
Drops
Type of Protection Against Electrical Shock
Serial Numbers 850000 or greater:
Type of Protection Against Electrical Shock
Protection Against Access to Hazardous Parts
with a Finger and Vertically Falling Water Drops
All Serial Numbers:
Expected Battery Life
Expected Life of Device
When integrity of external Mains is in question - use battery power only
Mode of Operation
Operating Temperature
Storage Temperature
Sound Pressure Level
Degree Of Protection Against Electrical Shock
Operating Barometric Pressure
IPX1
Class I
Class II
IP21
2 years (200 full charges)
5 years
Continuous
5°C - 40°C (41°F - 104°F)
95% R.H. non-condensing
0°C - 50°C (32°F - 122°F)
10% - 95% R.H. non-condensing
Does not exceed 70dB(A)
TYPE BF (Administration Sets are non-
conductive)
700 hPa (10.2 p.s.i.) - 1060 hPa (15.4 p.s.i).
Pump components are not affected by pres-
sure conditions.