To clear volume delivered:
1. Press RUN/PAUSE to stop pump.
2. Press RESET to clear Volume Delivered.
3. Press RUN/PAUSE to restart pump.
To view or clear accumulated volume:
1. Press RUN/PAUSE to stop pump.
2. Press and hold SET DOSE LIMIT for 5 seconds.
Total VOL and accumulated volume will be
displayed for 10 seconds.
3. To clear accumulated volume, press RESET
within the 10 seconds.
4. Press RUN/PAUSE to restart pump.
To view or change dose complete alarm mode:
Three alarm modes are available when dose is
complete:
• Mode 1 - Pump beeps then stops.
• Mode 2 - Pump beeps, stops and then beeps
at 5 minute intervals.
• Mode 3 - Pump beeps continuously.
To select one of the desired modes:
1. Press
and
for 5 seconds. The Infusion
Rate and Volume Delivered Displays should
show 888 and 8888 respectively.
2. Press
and
again. The Infusion Rate
Display will show 012 and the Volume
Delivered Display will show 3456.
3. Press
and
once again. The Infusion
Rate Display will show the current alarm
mode (1r, 2r or 3r) (Disregard the Volume
Delivered Display).
4. Select desired mode by pressing
5. Press RUN/PAUSE to save new alarm mode.
NOTE: If RUN/PAUSE is not pressed, new
alarm mode will not be saved.
6. Turn pump off.
7. Pump is now set to operate in the new
alarm mode.
Self Test
The major functions of the pump may be
tested by pressing the ON button and then
simultaneously holding the
three seconds.
The volume delivered and rate displays should
show all eights (8888) and the pump rotor
should turn. To test the alarms, press SET
DOSE LIMIT,
,
, RESET or RUN/PAUSE.
When any of these keys are pressed or the
flow sensors detect a drop, all alarm lights
should light, the alarm should sound and the
rotor will halt momentarily.
Any failure indicates a service check should be
performed. See Service Section for instruc-
tions on reporting a pump in need of service.
Service
EntraFlo™ Nutrition Feeding Pumps are reliab-
le medical electromechanical devices. As with
any mechanical device, problems may occur.
The following tips will help you correct the
minor in-service problems that may arise.
Note: Upon receipt of a new pump, plug it in
for a minimum of 12 hours to charge the back-
or
.
up battery. For maximum battery life, charge
the battery at least every 30 days, even if
the pump is not in use. Whenever the backup
battery is used to power the pump, charge for
12 hours as soon as conveniently possible.
and
keys for
EN
5