Current Measurement

Filter Function

FILTER OFF
Measured value including noise
Frequency characteristics when using the
filter function (100 A input)
Frequency [Hz]
20
Press for 1 sec.
FILTER OFF
FILTER ON Passband 100 [Hz]
FILTER ON
Measured value with reduced noise
Turn off the filter function when performing
measurement of power supply frequencies in excess
of 100 Hz, for example on an aircraft or ship.
Filter OFF