Ise Calibration - wtw pH/Ion 340i Instrucciones De Operación

Medidor manual del ph/ion
Idiomas disponibles

Idiomas disponibles

ISE calibration

ConCal (ISE)
Calibration
ISE
1
0 0
0
S
t
d
1
ConCal
ISE
2
0 0
0
t
2
S
d
ConCal
ISE
S
2
4
6
mV
10
ISE calibration
This is the conventional two-point or three-point calibration procedure that
uses 2 or 3 freely selectable standard solutions.
Use two or three different standard solutions from the following selection. In
the three-point calibration, the standards must be selected in either ascend-
ing or descending order.
Parameter
Std 1
Std 2
Std 3
Note
If a printer is connected to the interface, a calibration protocol is automatically
printed after a valid calibration.
ConCal (ISE) procedure:
l Press the <CAL> key. Std2 appears on the display.
l Enter the current concentration of the first standard solution with
< > < >.
mg/l
l Immerse the ISE electrode and, if necessary, the temperature sensor in
the first standard solution.
l Press <RUN/ENTER>. The AutoRead measurement begins. The elec-
trode voltage (mV) appears on the display. If the measured value is stable,
Std2 appears.
l Enter the current concentration of the second standard solution with
< > < >.
l Thoroughly rinse the ISE electrode and, if necessary, the temperature
mg/l
sensor with deionized water and immerse in the second standard solution.
l Press <RUN/ENTER>. The AutoRead measurement begins. The elec-
trode voltage (mV) appears on the display. If the measured value is stable,
the slope appears in calibration range I.
Note
At this point, the ISE calibration can be terminated with <M>. This
corresponds to a two-point calibration with the displayed slope.
Values [mg/l]
0.01 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1 / 2 / 5 / 10 / 20 /
50 / 100 / 200 / 500 / 1000
>
If Std 2
Std 1, Std 3 must be
If Std 2
<
Std 1, Std 3 must be
pH/Ion 340i
>
Std 2
<
Std 2
loading