• The test plug and test plug components may be opened or
repaired only by authorized Ottobock Service technicians.
• No liability will be assumed if the test plug is operated incorrectly
or used for purposes other than as originally intended.
3 Operation
For testing of a component proceed as follows::
1. Put all switches of the 757T16 System Tester in mid-position.
2. Connect the accumulator via appropriate cable to the 'Battery' receptacle
(Fig. 1). Important for the test: Use fully charged batteries!
3. Connect 757P28 Test Plug to the receptacle marked with 'Terminal
Device' (Fig. 1) and to the component to be tested.
4. Keep the enabling button pressed while pressing one of both function
keys simultaneously: Hand/Greifer should open or close (Fig. 2).
3.1 LED is lit (Fig. 3)
• The switch of the test plug is on.
• The maximum gripping speed and grip force are generated during the
functional test run.
3.2 LED is off (Fig. 4)
• Test plug switch is off.
• System Electric Hand/Electric Greifer close with minimum gripping speed
and grip force.
To ensure that the tests of the MYOBOCK components are performed
correctly and above all to be able to apply the results, it is necessary
to attend to a MYOBOCK Service Seminar at Ottobock that is offered
upon request.
4 Disposal
These products may not be disposed of with household waste
in some jurisdictions. Disposal that is not in accordance with
the regulations of your country may have a detrimental impact
on health and the environment. Please observe the information
provided by the responsible authorities in your country regarding
return and collection processes.
Ottobock | 7