6. Operation

1
Switching on the lamp
 To connect to the mains, fully insert the plug into the socket. Once inserted, the „5 min" LED will
flash. This is the readiness indicator (standby). Press the On/Off button (4) once to switch on the lamp.
2
Enjoying the light
While using the lamp on the face, always keep the eyes closed in order to avoid injury to the eyes (do
not look directly at the infrared light).
The recommended duration of treatment is maximum 15 minutes and should be limited to a ma-
ximum of two treatments a day. We recommend selecting a shorter radiation period to start with.
However, the skin of some people may show an over-sensitive (e.g. with extreme redness, blistering,
itching, heavy sweating) or allergic reaction to heat radiation, even when the lamp is used correctly.
3
LED display/timer
The IL 35 infrared lamp indicates your current treatment time with the aid of 3 LEDs.
The treatment time can be set at the following 3 levels.
To scroll through the timer, hold down the On/Off button (4) until the required timer level lights up.
The levels are as follows:
Treatment time
5 minutes
10 minutes
15 minutes
The lamp switches off automatically once the set treatment time has elapsed. To switch off the lamp
manually, repeatedly press the On/Off button (4) until only the „5 min" LED is flashing, the other LEDs
are off and the lamp goes out.
4

Important instructions

The distance between the infrared lamp and the radiated part of the body should not be less than the
following: at least 32 in (80 cm)
Basically,
examine the radiated section of the body regularly. Discontinue radiation immediately if any signs of
an over-sensitive or allergic reaction appear and consult a physician.
5
Switching off the lamp
CAUTION!
The lamp is hot after being used. Let the lamp cool off first long enough before you put it
away and/or pack it up!
Please note that the mains cable must not be wound round the infrared lamp either for storage or
when in use.
Disconnect the mains plug from the socket to fully switch off the device.

7. Changing the infrared lamp

The device must be switched off and allowed to cool and the mains plug pulled out before cleaning and chan-
ging the infrared heat lamp.
Please note that only infrared lamps of the original type may be used as replacements.
Replacement infrared lamp (type)
Infrared PAR 38 (150 W)
Warning
• Do not touch the infrared lamp or unscrew it when it is connected to the mains.
• Unplug the device, to avoid the risk of electrocution.
• In the interest of environmental protection, do not dispose of the device in household waste.
Number of lit LEDs
1
2
3
:
: Item number:
Specifications
230V ~ 50 Hz
10
616.51