RS-232 CONTROL
Interface:
Baud Rate:
Remote Functions:
Setting Accuracy:
Setting Resolution:
Readback Accuracy:
Readback Resolution:
GENERAL
Output Terminals:
AC Input:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Safety:
EMC:
Size:
Weight:
RS-232 interface, fully opto-isolated from power supply output.
9-pin D connector.
Variable from 600 baud to 9,600 baud.
Set Voltage, Set Current, Set Output On/Off, Read Voltage,
Read Current, Read On/Off, Read Mode (CV or CC).
Voltage 0.3% ± 20mV. Current 0.6% ± 20mA.
Voltage 10mV. Current 10mA.
Voltage 0.3% ± 100mV. Current 0.6% ± 20mA.
Voltage 100mV. Current 10mA.
4mm 'binding post' terminals on 19mm (0.75") pitch.
230V AC or 115VAC ± 10%, 50/60Hz. Installation Category II.
160VA max.
+5ºC to +40ºC, 20% to 80% RH.
−40ºC to +70ºC.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Complies with EN61010-1.
Complies with EN61326.
140 x 160 x 295 mm (WxHxD)
4.1kg.
3