Faro FS1010 Istrucciones De Uso página 12

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

 Before cleaning: cut the sealer from current! Clean the sealer only with a dry or
slightly moistened smooth wipe. Lever let water come into the sealer.
 Only use the sealer in dry rooms at ambient temperatures between 15 and 30°C. If the
sealer is used under conditions out of this specification, functional errors could occur
or the sealer could be damaged. Let the sealer come to room temperature before taking
it into operation.
 Only seal suitable packaging material compliant with EN 868-5 and ISO 11140-1. In
doubt contact your dealer or the manufacturer of the sealer.
 Never seal polyethylene pouches or similar. This will damage your sealer.
 Never feed pointed or flat instruments into the in feed gap. This could damage the
sealer or the instruments.
 Never insert parts into the ventilation gaps of the sealer. You could be injured by an
electric shock or the sealer could be damaged.
 Check electrical safety of the sealer at least every year: electric strength of insulation,
current leakage and ground connection resistance.
 Never use the sealer if there is any doubt if it is in a safe condition.
 The user is responsible for a acceptable sealing result. Let the seam be checked
frequently. Let the sealing process be validated.
Istruzioni d'uso/Instructions for use/Mode d'emploi/Istrucciones de uso
Automatic Sealer/Sigillatrice automatica FS 1010 – Rev.00
12
Tabla de contenido

Solución de problemas

loading

Productos relacionados para Faro FS1010

Tabla de contenido