Roche Diagnostics Accu-Chek Performa Nano Instrucciones De Uso página 31

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

7
or
(for example)
The control result, the
Press
bottle symbol, and a
control result as a
flashing L appear on the
Level 1. Press
display. Do not remove
second time to mark the
the test strip yet.
control result as a
Level 2.
53074_07214103002_EN.indb 29
Control Tests
to mark the
Press and release
meter.
a
OK and the control result alternate on the display if
the control result is in range. Err and the control
result alternate on the display if the control result is
not in range.
Remove and discard the used test strip.
After a successful test, the meter turns off
5 seconds after the test strip is removed.
4
8
or
(for example)
to set the control level in the