Digital field display
Effect
Under range. An input signal of <
Display
3.6 mA is connected to unit input.
indicates:
„
"
Over range. An input signal of >
Display
21.0 mA is connected to unit input.
indicates:
„
"
Operating parameters have been
Display
changed. The unit requests a save
indicates:
command.
„
"
The unit is saving changes made to
Display
the operating parameters.
flashes:
„
"
Effect
There is no circuit connected.
No measu-
red value
display
Unit defective.
Display defective.
Measured input incorrectly set up.
The 7 seg-
ment dis-
play shows
Sensor calibration faulty.
an incor-
rect/inac-
curate mea-
Offset incorrectly set up.
sured value
10.3
System messages
Cause
10.4 Fault analysis
Cause
10 Fault-finding
Cure
Please check the input signal.
Please check the input signal.
Release save/do not save using the "+/-"
push buttons and acknowledge with "E".
Once the save sequence has been comple-
ted the unit returns to displaying the mea-
sured values.
Cure
Please check the wiring to the unit.
Please exchange the unit.
Please exchange the unit.
Change set up in operating menu.
Compensate error using the offset.
Check offset.
19