Code
Cause
Text mes-
sage
•
S601
Danger of ves-
sel overfilling
Overfilling
Maintenance
The following table shows the error codes and
text messages in the status message "Mainte-
nance" and provides information on causes as
well as corrective measures.
Code
Cause
Text mes-
sage
•
M500
The data could
not be restored
Error during
during the reset
the reset "de-
to delivery
livery status"
status
•
M501
Hardware error
EEPROM
Error in the
non-active li-
nearisation
table
•
M502
Hardware error
EEPROM
Error in the
diagnostics
memory
•
M503
The echo/noise
ratio is too small
Measu-
for reliable
rement
measurement
reliability too
low
•
M504
Hardware defect
Error at a
device inter-
face
•
M505
Level echo can
no longer be
No echo
detected
available
VEGAPULS WL S 61 • 4 ... 20 mA
8.4
Rectification
Reaction when malfunction occurs
The operator of the system is responsible for
•
Make sure
taking suitable measures to rectify faults.
that there is no
further filling
Procedure for fault rectification
•
Check level in
The first measures are:
the vessel
•
•
•
Further comprehensive diagnostics options
are available with a PC with PACTware and the
suitable DTM. In many cases, the reasons can
be determined in this way and faults rectified.
Rectification
Check the 4 ... 20 mA signal
Connect a multimeter in the suitable measuring
•
Repeat reset
range according to the wiring plan. The following
•
Load XML file
table describes possible errors in the current
with sensor data
signal and helps to eliminate them:
into the sensor
•
Send instrument
for repair
•
Send instrument
for repair
•
Check installa-
tion and process
conditions
•
Clean the
antenna
•
Change polari-
sation direction
•
Check connec-
tions
•
Send instrument
for repair
•
Clean the
antenna
•
Remove possi-
ble false echoes
•
Optimize sensor
position and
Reaction after fault rectification
orientation
Depending on the reason for the fault and the
measures taken, the steps described in chapter
"Setup" must be carried out again or must be
8 Diagnosis, asset management and service
Rectify faults
Evaluation of fault messages via the adjust-
ment device
Checking the output signal
Treatment of measurement errors
Error
Cause
•
4 ... 20 mA
Fluctu-
signal not
ations
stable
of the
measured
variable
•
4 ... 20 mA
Electrical
signal mis-
connec-
sing
tion faulty
•
Voltage
supply
missing
•
Operating
voltage
too low
or load
resistance
too high
•
Current sig-
Elect-
nal greater
ronics
than 22 mA
module in
or less than
the sensor
3.6 mA
defective
Rectification
•
Set damping approp-
riate to the instrument
via the display and
adjustment module or
PACTware/DTM
•
Check connection
according to chapter
"Connection steps" and
if necessary, correct
according to chapter
"Wiring plan"
•
Check cables for
breaks; repair if neces-
sary
•
Check, adapt if neces-
sary
•
Exchange the instru-
ment or send it in for
repair
33