2-3 Electrical Specifications
Accuracy is ± (% reading + number of digits) at
23 C ± 5 C less than 75% R.H.
The above specifications do not include thermocouple error.
Measurement
Type
Accuracy
Range
-200.0 C ~ -100 C
± (0.3% reading + 1.0°C)
± (0.3% reading + 2.0 F)
-328.0°F ~ -148.0°F
-99.9 C ~ 999.9 C
± (0.1% reading + 0.7°C)
K
± (0.1% reading + 1.4 F)
-147.8°F ~ 999.9°F
1000 C ~ 1372 C
± (0.3% reading + 1°C)
± (0.3% reading + 2 F)
1000°F ~ 2501°F
-210.0 C ~ -100.0 C
± (0.3% reading + 1.1°C)
± (0.3% reading + 2.2 F)
-346.0°F ~ -148.0°F
-99.9 C ~ 999.9 C
± (0.1% reading + 0.8°C)
J
± (0.1% reading + 1.6 F)
-147.8°F ~ 999.9°F
1000 C ~ 1200 C
± (0.3% reading + 1°C)
± (0.3% reading + 2 F)
1000°F ~ 2192°F
T1 - T2 (K-, J-type), 55II only
± (0.3% of T1-T2 reading + 2.0 C)
± (0.3% of T1-T2 reading + 4.0 F)
Input Protection: 60V d.c. or 24V r.m.s.
Resolution: 0.1°C / 0.1°K / 0.2°F ≦ 1000°
1°C / 1K / 2°F ≧ 1000°