Standard operations
Calibration
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
Calibration procedure
This procedure is for general use with liquid calibration solutions. Refer to
the documents that are included with each probe for additional information.
Note: Solutions should be stirred during calibration. For more information about the
stirring settings, refer to
Change the stirring settings
1. Pour the buffers or calibration solutions into the labeled calibration
beakers.
2. From the main menu push
3. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
4. Push
to measure the first calibration solution.
The next calibration solution is shown.
5. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
6. Push
to measure the second calibration solution.
The next calibration solution is shown.
7. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
8. Push
to measure the third calibration solution.
When the calibration is good, the display briefly shows 3rd Buffer OK
and then returns to the main menu.
View the calibration data
Data from the most recent calibration can be shown.
8 English
on page 9.
to select pH or mV calibration.
1. From the main menu push
2. Use the
to view the last calibration data. The meter returns
automatically to the measurement screen after 3 seconds.
Set the calibration reminder
The calibration reminder can be set between 0 to 23 hours or 1-7 days.
The display shows the remaining time to the new calibration.
1. From the main menu push
Cal. frequency.
2. Use the
or
to change the value.
The meter returns automatically to the measurement screen after 3
seconds.
Sample measurements
Each probe has specific preparation steps and procedures for taking
sample measurements. For step-by-step instructions, refer to the
documents that are included with the probe.
Note: Solutions should be stirred during measurement. For more information about
the stirring settings, refer to
Change the stirring settings
Note: The temperature is not shown during an ORP (mV) measurement when the
temperature probe is not connected.
Push pH or ORP (mV) to take a sample measurement. During
measurements, the parameter flashes and the timer shows the
stabilization time. To change the measurement parameter (if applicable),
push and hold pH or ORP (mV).
To measure the sample continuously, push pH or ORP (mV) to start a
measurement and push pH or ORP (mV) again during the stabilization.
The parameter flashes to indicate the continuous measurement mode.
Advanced operations
Change the date and time
The date and time can be changed from the Date and Time menu.
1. Use the
to enter the Date and Time menu. The date and time format:
dd-mm-yyyy 24h.
.
and then push
until the display shows
on page 9.