During heating the selected temperature can be displayed for 1 second
by pressing MODE.
•
When the selected temperature has been reached, the heater demagnetises the
workpiece, switches off, and generates an acoustic signal for 10 seconds or until
START / STOP is pressed.
•
Press START / STOP to stop the heater.
•
Remove the workpiece with proper handling equipment.
•
If the workpiece remains on the heater, the heater will start again when the
temperature of the workpiece drops 10 °C (18 °F). Press START / STOP to stop the
heater and demagnetise the workpiece.
•
The TIH 030m is now ready to heat another workpiece with the same settings.
5.4 Time mode
•
If the remote control display shows °C or °F, press MODE to select TIME MODE.
The remote control display shows 't' in TIME MODE.
•
Press UP or DOWN to adjust the time in steps of 0.1 minute.
•
Press POWER to select the power level. Use the guidelines in section 5.8 to
determine the correct power setting.
•
Press START / STOP to start the heater. The remote control display shows the time
that remains.
•
During heating the temperature measured by the probe can be displayed for a couple
of seconds by pressing MODE.
•
When the time has elapsed, the heater demagnetises the workpiece, switches off, and
generates an acoustic signal for 10 seconds.
•
Press START / STOP to cancel the acoustic signal and stop the heater.
•
Remove the workpiece with proper handling equipment.
•
The TIH 030m is now ready to heat another workpiece with the same settings.
5.5 Temperature measurement
When the heater is not operating, the temperature of the workpiece can be measured by
pressing MODE and START / STOP at the same time.
The LED on the START / STOP button lashes during temperature measurement.
Press START / STOP to cancel temperature measurement.
5.6 Change of temperature unit
Press MODE and UP at the same time to switch between °C and °F.
The temperature unit setting remains the same even after disconnection from
mains power.