Technical Data - ProMinent DULCOTEST OZE 3-mA-2 ppm Instrucciones De Servicio

Ocultar thumbs Ver también para DULCOTEST OZE 3-mA-2 ppm:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Technical data

15 Technical data
Measured
variable
Ozone (O
Area of
application
Swimming pool, drinking water, industrial and process water or water
of similar quality (surfactant-free)
Measuring
ranges
OZE 3-mA-2 ppm:
Resolution
corresponds to the lower measuring range limit
Nominal slope
6 mA/ppm
Response time
T
90
T
90
pH range
Stability range of ozone
Temperature
range
5 - 45 °C, temperature-compensated
no temperature shifts
Pressure
DGM:
DLG III:
Flow rate
In-line probe DLG III or DGM
Cross-sensitivity
to chlorine < 2 %; chlorine dioxide interferes with the measurement;
surface-relaxing substances (e.g. surfactants, oils) block the diaphragm
(incorrect measurement)
Service life
Diaphragm cap
typically 1 year, depending on the water quality
If surface-relaxing substances (surfactants) are present, the service life may
be considerably reduced.
Service life
Electrolyte
typically 6 ... 8 weeks, depending on the water quality
Materials
Diaphragm cap:
Electrode shaft:
Supply voltage
16-24 Vdc, min. 35 mA at 16 Vdc
Output signal
4 - 20 mA
System of
protection
IP 65
Storage
temperature
5 - 50 °C
40
)
3
0.02 - 2 mg/l
(up):
approx. 180 s
(down): approx. 240 s
max. 1 bar (free drain!)
max. 1 bar (free drain!)
PVC transparent
PVC black and PMMA colourless
optimum:
30 l/h
minimum:
20 l/h
maximum: 100 l/h
®
ProMinent