ENGLISH
12
ASSEMBLY THE CUFF
1.Insert the edge of the cuff approximately 5 centimeters into the D-ring as shown
in figure.
2.Put the cuff on the left upper arm with the tube pointing to the direction of palm.
If measurement on your left arm is difficult, you can use right arm for measurement.
In this case, it is necessary to know that the readings may differ about 5-10
mmHg between left arm and right arm.
3.Wrap cuff around your upper arm with the lower edge of the cuff approximately
2-3 centimeters above the elbow. The mark <ART ERY > must be over the artery
of the arm.
4.Press the cuff to make sure that it is attached securely. The cuff should not be
too tight or too loose. Two fingers should be easily put in between cuff and upper
arm.
5.The mark <INDEX> on the cuff must point to area <NORMAL> (22-32 cm).
This means the cuff size is correct. If mark <INDEX> points to the area beyond
area <NORMAL>, please consult your dealer whether you need another size
cuff.
6.Sometimes it is difficult to make the cuff regular owning to the shape of the
user's upper arm, the cone-shape assembly of cuff is also acceptable.
7.If your clothes restrict blood circulation of your upper arm, or you roll your
sleeve up so as to result in such restriction. Please take off your clothes to get
accurate measurement if necessary.
ASSEMBLY THE STORAGE CASE
1.Three hooks of storage case aim at the concaves of device respectively;
2.Push the storage case upwards;
3.To fill tightly with the plug.
SETTING THE DATE AND TIME
The function provides accurate measuring time for each measurement.
To get accurate date and time, the user should preset the date and time correctly before the first use of this
device.