MAINTENANCE :
1 - The fibre-optic tip
After each patient use, the fibre-optic tip should be processed by disinfection and/or
sterilization. It is autoclaveable up to 135ºC.
2 - Any dirt or stains on the control unit or surface of the handpiece use a cloth wipe or
neutral cleansing.disinfection agent to clean.
NOTE:
Never use a solvent or volatile oil to for cleaning.
This will cause surface label printing to fade.
TECHNICAL DATA :
Main supply voltage:
Effect:
Classification:
Protection against electric shock:
Protection against harmful ingress of water: Ordinary equipment, IPX0
Mode of operation:
TROUBLE SHOOTING: (Guideline for fault diagnosis) :
PROBLEM
No power even when switch is on
No power even when switch is on
No light from Handpiece
Unexpected "beeping"
No light after a few uses (if bulb okay)
** In case of more serious malfunctions, please return to manufacture for repair.
100-240Vac, 50/60Hz
A
Class I
Type B applied part
Continuous operation or 30seconds
CAUSE
Cord or plug connections
Fuse
Halogen bulb not working
Some buttons are pressed
Cooling fan not working
METHOD
Inspect the connection of power
Cable plug and co-axial cable
Inspect fuse and change if needs,
If it fails again, please send for repair
Replace bulb
Check all buttons
Send for check and repair