External surface of the micro motor may be cleaned with a damp cloth or disinfectant wipe. Sterilize micro
motor with a steam sterilizer. Do not place micro motor in ultrasonic sterilizers or submerge it in any solutions.
a. Cleaning and disinfection
Motor may be cleaned with a damp cloth or disinfectant wipe.
b. Sterilization
- Autoclave: Class B.
- Sterilization temperature: 134°C (273°F)//2.1-2.2 bars.
- Time at sterilization temperature: 5 minutes
When treating patients who may have an acute, critical infectious disease, be sure to observe the
hygienic measures cited in applicable publications and reports. If possible, use suitable disposable
products to avoid the transmission of critical pathogens. These protect the user, the patient and all participants
in the surgery. A microbiologically appropriate coolant liquid must be used. Only use prescribed isotonic saline
solution NaCl 0.9 (identified as an infusion solution; follow the instructions on the package insert) for cooling
and spraying wounds.
Not all autoclaves can reach 134° C. Not all autoclaves draw a pre-vacuum. Please refer to
your autoclave manufacturer for specific sterilization instructions.
XI – MONITORING / PREVENTIVE AND CORRECTIVE MAINTENANCE
Important:
In the case of a fault, you are advised to contact the supplier of your device. Do not use unauthorized
repairers, who might make your device dangerous for you and your patients.

11.1 MONITORING

Regular monitoring of the device and its accessories is necessary in order to detect any insulation fault or
damage. Replace, if necessary.
It is important to keep the control unit ventilation vents clean in order to avoid abnormal heating.

11.2 PREVENTIVE AND CORRECTIVE MAINTENANCE

I-SURGE + micro-motor inspection/overhaul: The manufacturer recommends inspection or overhaul of the I-
SURGE + micro-motor at least once a year.
The micro-motor cable has limited durability and is considered a consumable product. The durability of the cable
is largely dependent upon use and treatment, i.e., handling, reprocessing and frequency of use. Caution must be
exercised when connecting or disconnecting the micro motor and foot control cable from the device to avoid
damage to the cable (Fig. 8). Motor assembly has been tested for 500 sterilization cycles.
Periodically inspect the micro motor cord for damage prior to use. If there is damage noted, contact SATELEC.
1. Do not attempt to disassemble the motor or motor connector.
2. Do not oil or lubricate the motor.
3. Do not attach a handpiece to the motor while the motor is running.
4. Do not bend motor cord sharply.
5. Do not drop or impact micro motor against hard surface.
Failure to comply with any of the above instructions may avoid your warranty