In every program Programs [1 to 5] are set the same parameters by the manufacturer:
Ratio : 20:1
Speed : 2000 rpm
Flow : 3 (Max)
Torque : maxNcm
Motor sense: FWD
The listed parameters are only examples. To prevent unnecessary risk, observe the guideline
speeds given by the manufacturer of the rotating instruments and the guideline given by the
manufacturer of the "implant System" in use.

7.2 SETTINGS

I-SURGE will store the operative parameters for each "program" as soon as they are set by the user. They will be
maintained in the memory even if the central unit is switched off. For each "program" it is possible to set:
speed, torque, ratio, direction of rotation and flow. To reset the parameters of the manufacturer (Software Unit
Reset), press the "PUMP ON/OFF" key while the unit is switched on (Fig. 6).

7.3 TORQUE CALIBRATION

Perform torque calibration, press "Calibration" key on front panel while holding the motor (not drilling). This
function checks the handpiece mechanical inertia, in order to off-set this value.
This operation will take a few seconds, the drill will run at different speed (Fig. 7).
The drill must not be engaged in any operation when this function is performed.
Calibration should be done before each surgical session or anytime the handpiece is
changed or lubricated.
Do not perform calibration if the micro motor is not inserted in the receptacle on the
front panel of the unit. In case of error, perform a "System Reset" as described
previously.

7.4 PROGRAM SETTINGS ADJUSTMENT

a. Contra-angle selection
RATIO: Sets handpiece reduction ratio. See the available reduction ratio in the following table:
1:1
16:1
20:1
24:1
32:1
64:1
80:1