CAUTION
Use under unallowable environmental conditions
Risk of injury due to damage to the product
► Do not expose the product to unallowable environmental conditions.
► If the product has been exposed to unallowable environmental condi­
tions, check it for damage.
► If damage is apparent or in case of doubt, do not continue using the
product.
► Take suitable measures if required (e.g. cleaning, repair, replacement,
inspection by the manufacturer or a specialist workshop etc.).
CAUTION
Exceeding the service life and reuse on another patient
Risk of injury due to loss of functionality as well as damage to the product
► Ensure that the approved service life is not exceeded.
► Only use the product on a single patient.
CAUTION
Mechanical damage to the product
Risk of injury due to change in or loss of functionality
► Use caution when working with the product.
► If the product is damaged, check it for proper function and readiness
for use.
► In case of changes in or loss of functionality, do not continue using the
product (see "Signs of changes in or loss of functionality during use" in
this section).
► Take any necessary measures (e.g. repair, replacement, inspection by
the manufacturer's customer service, etc.).
Signs of changes in or loss of functionality during use
Among other factors, changes in functionality can be indicated by an altered
gait pattern, a change in the positioning of the prosthetic components relat­
ive to each other and by the development of noises.

4 Scope of delivery

Item
Quantity
1
2
4R153* Adapter plate (see fig. 1)
Designation
1
Adapter plate (with or without offset)
1
Torx offset screwdriver
Reference number
709S51=T8
13
Tabla de contenido
loading

Este manual también es adecuado para:

4r153-1

Tabla de contenido