3. Starting and stopping the device
This device should only be used by professionals (dental
practitioners and hygienists). Before being used for the
first time it is:
•
imperative to clean, disinfect using alcohol or
disinfectant dental wipes and/or sterilize the
accessories (casing of the device, handpiece, tips,
wrenches, etc.);
•
recommended that the accessories be plugged in and
the spray operated for a few minutes to rinse the
device's internal water circuits.
Important: It is imperative that the water supply is turned
off when leaving the dental practice.
Switching on the power
1.
Set the power switch to the on position (I) (rep.
30).
The device automatically goes into standby
mode: the green indicator
front lights up. The device is in energy saving
mode.
To operate
1.
Press the
(ref. 25) key on the front of the
device
The initialization phase lasting a few seconds is
launched. The information displayed is as follows
(any other values indicate malfunction):
Display: the figure 88 is displayed.
-
Background color of the display: bluish-
-
white.
Indicators: they all light up.
-
At the end of the initialization phase, the display
will show one of the following:
by default, low power mode is at level 5 on a
-
green background, no irrigation, water
supply selected, light out in three seconds,
(Prophy Max Lux), enters standby mode
automatically after nine minutes;
or the last configuration used.
-
The device is ready to use.
6
Entering standby mode
1.
Press the
-
The front face goes out.
-
The device is in energy saving mode.
The
-
Reminder: the device enters standby mode on its own after
a predetermined time. See § Setting the device's standby
delay.
Leaving standby mode
1.
Press the
The
(rep 10) indicator goes out and the device
returns to the configuration being used before
entering standby.
Changing the delays
(ref. 10) on the
Delays refer to the time the light in the Suprasson
handpiece stays on after ultrasound has been used, and the
time taken before the device goes into standby mode after
a period of inactivity.
Delay before the light goes out
This function only applies to the Prophy Max Lux. It
specifies the length of time before the light goes out after
ultrasound has been used (three seconds by default). To
change this setting:
1.
Put the device into standby (
2.
Press the yellow and boost keys together, then
the
depressed.
The display will indicate «L» (Light) followed by a
figure giving the current time delay in seconds.
3.
Release the keys.
4.
Set the time with the
The length of time can be set from 0 (L-) to nine
seconds (L9). The new value comes into effect
when the device is restarted.
Therefore a light out time delay of 7 seconds will be
displayed at "L7".
(rep 25) key
(rep 10) indicator lights up.
key (rep 25)
®
Lux
key).
key while keeping the other keys
or
keys.
Delay before the device enters standby mode
This setting specifies the length of time before the device
enters standby mode. To change this setting:
1.
Put the device into standby (
2.
Press the yellow and blue keys together, then
the
key, while keeping the other keys
pressed.
The display will indicate "A" (Stop) followed by a
figure giving the current time delay in minutes.
3.
Release the keys.
4.
Set the time with the
or
The length of time can be set from no standby to
15 minutes. The new value comes into effect
when the device is restarted.
"AC" indicates that standby mode will be activated after 12
minutes. The table below shows the value displayed
according to the desired time delay in minutes.
Display
min
Display
A-
0 (*)
A8
A1
1
A9
A2
2
AA
A3
3
AB
A4
4
AC
A5
5
AD
A6
6
AE
A7
7
AF
(*) 0 = no standby
Switching off the device
1.
Set the on/off switch to the 0 position.
key).
keys.
min
8
9
10
11
12
13
14
15