SAFETY TIPS AND GUIDELINES (CONTINUED)
CAUTION
S
Do not steam clean or hose off the bed. Do not immerse any part of the bed. The internal electric parts
may be damaged by exposure to water. Hand wash all surfaces of the bed with warm water and mild
detergent. Dry thoroughly. Inspect the mattress cover after each use. Discontinue use if any cracks or
rips are found in the cover which may allow fluids to enter the mattress. Exposure to fluids may cause
injury to patient and/or user.
S
Preventative maintenance should be performed at a minimum of annually to ensure all bed features are
functioning as designed. Close attention should be given to safety features including, but not limited to:
S Safety side latching mechanisms,
S Frayed electrical cords and components,
S Protective earth ground impedance 100 milliohms max.
S No controls or cabling entangled in bed mechanisms,
S All electrical controls return to off or neutral position when released,
S For additional maintenance information, refer to your maintenance manual.
S
Always unplug bed during service or cleaning. When working under the bed with the bed in the high posi-
tion, always place blocks under the litter frame and set the brakes to prevent injury in case the Bed Down
switch is accidently pressed.
S
Hand wash all surfaces of the bed with warm water and mild detergent. Dry thoroughly. DO NOT STEAM
CLEAN, PRESSURE WASH, HOSE OFF OR ULTRASONICALLY CLEAN. Using these methods of
cleaning is not recommended and may void this product's warranty.
Inspect the mattress cover after each use. Discontinue use if any cracks or rips are found in the cover
which may allow fluids to enter the mattress. Exposure to fluids may cause injury to patient and/or user.
S
To avoid injury, unplug the bed power cord from the wall socket before using the manual hand crank.
S
The LD304 Birthing Bed is equipped with a hospital grade plug for protection against electric shock haz-
ard. It must be plugged directly into a properly grounded three−prong receptacle. Grounding reliability
can be achieved only when a hospital grade receptacle is used.
S
To avoid damage, the weight of the IV bags should not exceed 40 pounds.
S
To avoid damage while transporting the bed, verify the IV pole is at a low enough height to allow it to pass
safely through door openings.
S
IV Poles should not be used as a bed push/pull device.
S
The cleanliness and integrity of both ground chains must be maintained to minimize static build−up and
discharge.
Introduction
S Caster braking system,
S Leakage current 300 microamperes max.
1−5