Blood Pressure Accuracy Check Method - Gima SPOT-CHECK PC-300 Guía De Uso

Ocultar thumbs Ver también para SPOT-CHECK PC-300:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Table 1 ECG Leads Configuration and Electrodes Location Table
Electrode
Name & Color
Electrode
Location
The intersection between the centerline
of the right clavicle and Rib 2.
The intersection between the centerline
of the left clavicle and Rib 2.
Between the left edge of the breast
bone and Rib 5.

2.4 Blood Pressure Accuracy Check Method

Operation procedure:
1. Unscrew the M3x6 screw from the battery compartment on the back of the Spot-Check Mo-
nitor, as shown in figure 2.11.
2. Take a NIBP connector plug from the battery cover, as shown in figure 2.12. (Note: there are
two plugs but you will only need one.)
3. Air Path Connection: Take a piece of air tube (0.5~1m long, Φ8.0mm/Φ4.0mm diameter). Atta-
ch the NIBP connector with a connector plug on to one end of the air tube. Connect the other
end to the 3-way connector. Connect the other 2 ends of the 3-way connector to a Mercury
Sphygmomanometer as shown in Figure 2.13.
ENGLISH
Lead
Lead I
Lead II
R (Red)/
R (Red)/
RA (White)
RA (White)
F (Green)/
L (Yellow)/
LL (Red)
LA (Black)
L (Yellow)/
F (Green)/
LA (Black)
LL (Red)
86
Lead III
L (Yellow)/
LA (Black)
R (Red)/
RA (White)
F (Green)/
LL (Red)
Tabla de contenido
loading

Este manual también es adecuado para:

35162

Tabla de contenido