Sartorius YTM03MA Instrucciones De Funcionamiento página 9

Kit ajuste de temperatura para ma45, ma50 y ma100
Ocultar thumbs Ver también para YTM03MA:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

§ Press the [
Start
temperature adjustment routine.
> The heating symbol appears in the display; the heat-
ing time and the previous drying time are displayed
continuously.
§ After you have finished the measurement for cali-
brating and adjusting the temperature, you will hear
3 beeps. Display: Analysis ended
§ Press [
Info
] to display the measured values.
§ Go back by pressing the , soft key
> Return to the main menu without performing an
adjustment by pressing the [
is printed out automatically.
§ Activate the calibration/adjustment and temperature
adjustment routine on the unit by pressing the
Adjust.
[
] soft key. This concludes the adjust-
ment routine. If the final temperature shows 163°C,
for example, the temperature is 1°C outside the
tolerance limit of 160°C ± 2°C. This adjustment
automatically corrects the deviation. At the same
time, a data record is printed out (for example,
on the Sartorius YDP02/03 data printer if you are
using the MA45).
] soft key to activate the
End
] soft key. A record
9
Tabla de contenido
loading

Tabla de contenido