Gima Sonoline C Manual Del Usuario página 23

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

23
Chapter 1
Safety Guidance
This unit is internally powered equipment and the degree of shock protection
is type CF applied part
connections will comply with permitted leakage currents, dielectric strengths
of IEC 60601-1.
WARNING and CAUTION messages must be observed. To avoid the possibility
of injury, observe the following precautions during the operation of the device.
WARNING: This device is not explosion-proof and can not be used in the
presence of flammable anaesthetics.
WARNING: Do not throw batteries in fire as this may cause them
to explode.
WARNING: Do not attempt to recharge normal dry-cell batteries, they may
leak, and may cause a fire or even explode.
WARNING: Don't touch signal input or output connector and the patient
simultaneously.
WARNING: Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC standards (e.g. IEC
950 for data processing equipment and IEC60601-1 for medical equipment).
Furthermore all configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional equipment to the
signal input connector or signal output connector configures a medical system,
and is therefore responsible that the system complies with the requirements
of the valid version of the system standard IEC 60601-1-1. If in doubt, consult
our technical service department or your local distributor.
WARNING: Pocket Fetal Doppler is a tool to aid the healthcare professional
and should not be used in place of normal fetal monitoring.
WARNING: Replacing battery shall only be done outside the patient
environment (1.5m away from the patient).
CAUTION: The device must be serviced only by authorized and qualified
personnel.
CAUTION: The main unit is designed for continuous operation and is
'ordinary'. Do not immerse in any liquid (i.e. not drip or splash-proof).
CAUTION: Keep the device clean. Avoid vibration.
CAUTION: Do not use high temperature sterilizing process and E-beam
or gamma radiation sterilization.
CAUTION: Electromagnetic Interference-Ensure that the environment
in which the device is operated is not subject to any sources of strong
electromagnetic interference, such as radio transmitters, mobile telephones, etc.
. Type CF protection means that these patient
ENGLISH
Tabla de contenido
loading

Este manual también es adecuado para:

29480

Tabla de contenido