6.2
Calibration of the meter
The Grease Meter can be calibrated for different grease types, characterisitics and
densities.
6.2.1
Calculate the correction factor
Correction factor = real quantity / shown quantity
Example:
500 grams of grease have been dispensed.
The Grease Meter shows 485 GRAM.
Correction factor = 500 / 485 = 1.03
6.2.2
Calculate the calibration factor
The new calibration factor =
the old calibration factor x the correction factor.
Example:
The old calibration factor is 700.
The correction factor is calculated as 1.03.
The new calibration factor = 700 x 1.03 = 721.
SKF LAGM 1000E
9