8.5

Changing applicators

To change the various applicators, hold the hand-piece in one hand and unscrew the applicator from
the hand-piece with the other hand (anticlockwise). Screw the required head tightly onto the hand-
piece (clockwise), until the black outside ring of the applicator rests on the hand-piece (there should
not be any thread visible).
NOTE:
Applicators are expendable parts and must be replaced after a certain period of use.
Minor / slight deformation or shortening of the rear impact dome does not affect functionality.
In cases of greater deformation or stronger shortening of the rear impact dome the applicator
must be replaced.
A test template is supplied with the device that enables the user to test if the wear limit has been
reached (see diagram).
Applicator measurement
tool
Air gap between applicator and he
applicator measurement tool
Applicator is OK
8.6

Foot switch

Place the foot switch so that it can be reached easily during treatment. The footswitch control unit is multi-
directional so it is not necessary to align the footswitch exactly.
To avoid damage, please note that only slight pressure needs to be put on the switch. Use the front part
of your foot, not the heel to operate the footswitch.
The switch does not have a locking device, which means that it only remains actuated as long as
pressure is applied.
Applicator measurement tool
Template makes contact or an air gap
at tip
Applicator need to be changed